Introduction

The E.Z.N.A.® Mollusc DNA Kit is designed for efficient recovery of genomic DNA up to 60 kb in size from molluscs, insects, arthropods, roundworms, flatworms, and other invertebrate tissue samples rich in mucopolysaccharides. The method is suitable for invertebrates frozen or preserved in alcohol or DNE solution, and good results can be obtained with formalin preserved material.

The procedure relies on the well established properties of the cationic detergent, cetyltrimethyl ammonium bromide (CTAB), in conjunction with the selective DNA binding of Omega Bio-tek’s HiBind® matrix. Samples are homogenized and lysed in a high salt buffer containing CTAB and extracted with chloroform to remove mucopolysaccharides. Binding conditions are adjusted and DNA further purified using HiBind® DNA spin columns. In this way, salts, proteins and other contaminants are removed to yield high-quality genomic DNA suitable for downstream applications such as endonuclease digestion, thermal cycle amplification, and hybridization techniques.

New in this Edition:

August 2019

• General edits. No change to the procedure.

October 2015

• MBL Buffer has been renamed BL Buffer. This is a name change only. The buffer formulation has not changed.

Kit Contents

Product Number

D3373-00

D3373-01

D3373-02

HiBind® DNA Mini Columns

5

50

200

2 mL Collection Tubes

10

100

400

ML1 Buffer

3 mL

30 mL

125 mL

BL Buffer

5 mL

20 mL

80 mL

HBC Buffer

5 mL

25 mL

80 mL

Proteinase K Solution

150 µL

1.5 mL

6.0 mL

RNase A

55 μL

520 μL

2.1 mL

DNA Wash Buffer

2.5 mL

25 mL

3 x 25 mL

Elution Buffer

2 mL

30 mL

60 mL

User Manual

 

 

 

Storage and Stability

All of the E.Z.N.A.® Mollusc DNA Kit components are guaranteed for at least 12 months from the date of purchase when stored as follows. RNase A should be stored at 2-8°C. Proteinase K Solution can be stored at room temperature for up to 12 months. For longterm storage, store Proteinase K Solution at 2-8°C. All remaining components should be stored at room temperature. During shipment or storage in cool ambient conditions, precipitates may form in some buffers. Dissolve such deposits by warming the solution at 37°C and gently shaking.

Determination of DNA Quality and Quantity

Dilute a portion of the eluted material approximately 10-20 fold in DNA Elution Buffer or 10 mM Tris, pH 8.5. Measure absorbance at 280 nm and at 260 nm to determine the A260/ A280 ratio. Values of 1.7-1.9 generally indicate 85%-90% purity. The concentration of DNA eluted can be determined as follows:

Concentration = 50 μg/mL x Absorbance260 x {Dilution Factor}

 

Preparing Reagents

1. Dilute DNA Wash Buffer with 100% ethanol as follows and store at room temperature.

Kit

100% Ethanol to be Added

D3373-00

10 mL

D3373-01

100 mL

D3373-02

100 mL per bottle

2. Dilute HBC Buffer with 100% isopropanol as follows and store at room temperature.

Kit

100% Isopropanol to be Added

D3373-00

1.6 mL

D3373-01

10 mL

D3373-02

32 mL

E.Z.N.A.® Mollusc DNA Kit Protocol

E.Z.N.A.® Mollusc DNA Protocol

Invertebrates preserved in formalin should be rinsed in xylene and then ethanol before processing. Note that results obtained with formalin-fixed tissues generally depend on age and size of specimen. Purified material is usually adequate for PCR amplification, but fresh or frozen samples should be used for Southern analyses.

Materials and Equipment to be Supplied by User:

• Microcentrifuge capable of at least 10,000 x g

• Nuclease-free 1.5 mL or 2 mL microcentrifuge tubes

• Water baths capable of 60°C and 70°C

• Vortexer

• 100% ethanol

• 100% isopropanol

• Chloroform:isoamyl alcohol (24:1)

• Optional: 3M NaOH for column equilibration

Before Starting:

• Prepare buffers according to the instructions on Page 4

• Set water baths to 60°C and 70°C

• Heat Elution Buffer to 70°C

1. Homogenize tissue sample following one of the procedures below depending on the sample type.

A. Arthropods

1. Pulverize no more than 50 mg of tissue in liquid nitrogen with a mortar and pestle.

Note: If ceramic mortar and pestle are not available, homogenize the sample in the microcentrifuge tube using a disposable microtube pestle (Omega Bio-tek, Cat No. SSI-1015-39; Eppendorf, Cat No. 0030 120.973; VWR, Cat No. KT 749520- 0000).

2. Transfer the powder to a clean 1.5 mL microcentrifuge tube.

3. Proceed to Step 2 below.

B. Molluscs (and other soft tissue invertebrates)

1. Pulverize no more than 30 mg tissue in liquid nitrogen with a mortar and pestle.

2. Transfer the powder to a clean 1.5 mL microcentrifuge tube.

Note: If ceramic mortar and pestle are not available, homogenize the sample in the microcentrifuge tube using a disposable microtube pestle (Omega Bio-tek, Cat No. SSI-1015-39; Eppendorf, Cat No. 0030 120.973; VWR, Cat No. KT 749520-0000). Addition of a pinch of white quartz sand, -50 to 70 mesh (Sigma Chemical Co. Cat No. S9887) will help.

3. Proceed with Step 2 below.

Note: The amount of starting material depends on the sample and can be increased if acceptable results are obtained with the suggested 30 mg tissue. For easy-to-process specimens, the procedure may be scaled up and the buffer volumes used increased in proportion. In any event, use no more than 50 mg tissue per HiBind® DNA Mini Column as binding capacity (100 μg) may be exceeded. Meanwhile, difficult tissues may require starting with less than 30 mg tissue and doubling all volumes to ensure adequate lysis.

2. Add 350 μL ML1 Buffer and 25 μL Proteinase K Solution. Vortex to mix thoroughly.

3. Incubate at 60°C for a minimum of 30 minutes or until entire sample is solubilized.

Note: Actual incubation time varies and depends on the elasticity of the tissue. Most samples require no more than 4 hours. Alternatively an overnight incubation at 37°C will produce adequate results.

4. Add 350 μL chloroform:isoamyl alcohol (24:1). Vortex to mix thoroughly.

5. Centrifuge 10,000 x g for 2 minutes at room temperature.

6. Transfer the upper aqueous phase to a clean 1.5 mL microcentrifuge tube. Avoid the milky interface containing contaminants and inhibitors.

Note: This step will remove much of the polysaccharides and proteins from solution and improve spin-column performance downstream. If a small upper aqueous phase is present after centrifugation, add 200 μL ML1 Buffer and vortex to mix thoroughly. Repeat Step 5 (centrifugation) and Step 6 (transfer the upper aqueous phase).

7. Add one volume BL Buffer and 10 μL RNase A. Vortex at maximum speed for 15 seconds.

Note: For example, to 350 μL upper aqueous solution from Step 6, add 350 μL BL Buffer.

8. Incubate at 70°C for 10 minutes.

9. Cool the sample to room temperature.

10. Add one volume 100% ethanol. Vortex at maximum speed for 15 seconds.

Note: For example, to 350 μL upper aqueous solution from Step 6, add 350 μL 100% ethanol.

11. Insert a HiBind® DNA Mini Column into a 2 mL Collection Tube. Optional Protocol for Column Equilibration: 1. Add 100 µL 3M NaOH to the HiBind® DNA Mini Column. 2. Centrifuge at maximum speed for 60 seconds. 3. Discard the filtrate and reuse the collection tube.

12. Transfer 750 µL sample from Step 9 (including any precipitate that may have formed) to the HiBind® DNA Mini Column.

13. Centrifuge at 10,000 x g for 1 minute.

14. Discard the filtrate and reuse the collection tube.

15. Repeat Steps 12-14 until all of the sample has been applied to the HiBind® DNA Mini Column.

16. Discard the filtrate and the Collection Tube.

17. Insert the HiBind® DNA Mini Column into a new 2 mL Collection Tube.

18. Add 500 μL HBC Buffer.

Note: HBC Buffer must be diluted with 100% isopropanol before use. Please see Page 4 for instructions.

19. Centrifuge at 10,000 x g for 30 seconds.

20. Discard the filtrate and reuse the Collection Tube.

21. Add 700 μL DNA Wash Buffer.

Note: DNA Wash Buffer must be diluted with 100% ethanol before use. Please see Page 4 for instructions.

22. Centrifuge at 10,000 x g for 1 minute.

23. Discard the filtrate and reuse the Collection Tube.

24. Repeat Steps 21-23 for a second DNA Wash Buffer wash step.

25. Centrifuge the empty HiBind® DNA Mini Column at maximum speed for 2 minutes to dry the membrane.

Note: It is critical to remove any trace of ethanol that may otherwise interfere with downstream applications.

26. Transfer the HiBind® DNA Mini Column to a nuclease-free 1.5 or 2 mL microcentrifuge tube (not provided).

27. Add 50-100 μL Elution Buffer (or sterile deionized water) preheated to 70°C.

Note: Smaller elution volumes will increase DNA concentration but decrease yield. Elution volumes greater than 200 μL are not recommended.

28. Let sit at room temperature for 2 minutes. 29. Centrifuge at 10,000 x g for 1 minute. 30. Repeat Steps 27-29 for a second elution step.

Note: Any combination of the following steps can be used to help increase DNA yield.

• After adding the Elution Buffer, incubate the column for 5 minutes.

• Increase the elution volume.

• Repeat the elution step with fresh Elution Buffer (this may increase the yield, but decrease the concentration).

• Repeat the elution step using the eluate from the first elution (this may increase yield while maintaining elution volume).

31. Store DNA at -20°C.

Troubleshooting Guide

Please use this guide to troubleshoot any problems that may arise. For further assistance, please contact the technical support staff , toll free, at (800-832-8896).

Problem

Cause

Solution

Clogged Column

Incomplete lysis

Increase incubation time with ML1 Buffer/ Proteinase K Solution. An overnight incubation may be necessary.

Sample too large

Do not use greater than the recommended amount of starting material. For larger samples, divide into multiple tubes

Incomplete homogenization

Pulverize material as indicated in liquid nitrogen to obtain a fine powder.

Low DNA yield

Clogged column

See above

Poor elution

Repeat elution or increase elution volume. Incubate the column at 70°C for 5 minutes before centrifugation.

Poor binding to column

Follow the protocol closely when adjusting the binding conditions.

Improper washing

DNA Wash Buffer must be diluted with 100% ethanol before use.

Low DNA Yield or no DNA Yield

Extended centrifugation during elution step

Resin from the column may be present in eluate. Avoid centrifugation at speeds higher than specified. The material can be removed from the eluate by centrifugation; it will not interfere with PCR or restriction digests.

DNA washed of

Repeat the DNA isolation, be sure to thoroughly mix the sample with ML1 Buffer.

Trace protein contaminants remain

Following Step 19, wash column with a mixture of 300 μL BL Buffer and 300 μL ethanol before proceeding to Step 20.

No DNA eluted

Poor cell lysis

Increase incubation time with ML1 Buffer. An overnight incubation may be necessary.

Incomplete homogenization

Pulverize starting material as indicated in liquid nitrogen to obtain a fine powder.

100% ethanol was not added before adding sample to column

Before applying DNA sample to column, add BL Buffer and 100% ethanol.

Ethanol was not added to the DNA Wash Buffer

Dilute DNA Wash Buffer with the indicated volume of 100% ethanol before use.

Isopropanol was not added to the HBC Buffer

Dilute HBC Buffer with the indicated volume of isopropanol before use.

Ordering Information

The following components are available for purchase separately.

(Call Toll Free at 1-800-832-8896)

Product

Part Number

HiBind® DNA Mini Columns, 200 columns

DNACOL-02

RNase A, 5 mL

PD090

Proteinase K Solution

AC116

Elution Buffer, 100 mL

PDR048

DNA Wash Buffer, 100 mL

PS010


  

 

제품자료보기

상품문의

등록된 상품문의

상품문의가 없습니다.

반품/교환정보

셀젠바이오샵에서는 다음과 같은 기간 및 내용으로 상품에 대하여 교환, 반품, 환불을 보장하고 있으며, 상품의 반환에 의한

추가비용을 고객에게 부담시지키 않습니다. (, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야

합니다.)

 

::: 교환 및 반품이 가능한 경우:::

-상품을 공급 받으신 날로부터 7일 이내

  , 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.

-공급받으신 상품 및 용역의 내용이 표시, 광고 내용과 다르거나 다르게 이행된 경우
-

:::교환 및 반품이 불가능한 경우:::

-하자가 없는 상품을 택배사로부터 인도 받으신 날로부터 20일 이후.
- 고객님의 책임 있는 사유로 상품등이 멸실 또는 훼손된 경우.
- 포장을 개봉하였거나 포장이 훼손되어 상품가치가 현저히 감소한 경우
- 시간의 경과에 의하여 재판매가 곤란할 정도로 상품등의 가치가 현저히 감소한 경우
- 주문 당시 해외 발주 상품의 경우
-

주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)

통해 요청하시면 친절하게 처리해 드리겠습니다.

주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.
  • ㈜셀젠바이오테크, 대표자명: 김성빈, 사업자등록번호: 344-08-00018, 통신판매업종 제 2017-대전유성-0220 호
  • 대전광역시 대덕구 대화로69, 3층(대화동)
  • 대표전화: 042-824-9026 팩스: 042-824-9027
  • E-Mail: celgen-bio@celgen-bio.com
  • Copyright © 2015 Celgen-Bio.tech All rights reserved.